Get Support

+8801977426642

Fluke ULT800 Electrical Safety Analyzer

  • Home
  • Fluke ULT800 Electrical Safety Analyzer

Fluke ULT800 Electrical Safety Analyzer in Bangladesh

The Fluke ULT800 Electrical Safety Analyzer is an advanced, professional-grade instrument designed for comprehensive electrical safety testing of medical equipment. In Bangladesh, it is widely used by hospitals, clinics, diagnostic centers, biomedical engineering departments, and service providers to ensure that medical devices operate safely and comply with regulatory standards.

The ULT800 combines high-accuracy measurements with versatile testing capabilities, allowing biomedical engineers and technicians to perform routine inspections, preventive maintenance, and post-repair verification efficiently and accurately. These features make the ULT800 an ideal solution for ensuring patient safety, improving maintenance efficiency, and maintaining regulatory compliance in healthcare environments. Pico Labs Limited is a reliable technology provider in Bangladesh where you can explore a wide range of Biomedical Test Equipment at the best price. From hospitals and clinics to diagnostic centers and biomedical engineering departments, you can find professional analyzers, simulators, and test tools for accurate performance checking of medical devices like infusion pumps, defibrillators, patient monitors, incubators, and ventilators.

Example Technical Overview Table

The table below presents a simplified overview of typical functions and characteristics of the Fluke ULT800 Electrical Safety Analyzer. Actual specifications may vary depending on configuration.

Parameter / Function Typical Description
Ground Wire Resistance Measures continuity of protective earth conductor
Equipment Leakage Current Measures leakage from mains and enclosure
Applied Part / Patient Leakage Tests current through patient-connected parts
Test Modes Automated and manual sequences for various device types
Display High-resolution digital numeric readout
Power Source Mains-powered
Portability Portable, suitable for field and bench use
Operation Intuitive interface with menu-driven navigation
Application Environment Hospitals, clinics, labs, service centers
Typical Use Comprehensive electrical safety testing of medical devices

This table is for general understanding and not a formal datasheet.

Applications in Bangladesh

The Fluke ULT800 Electrical Safety Analyzer serves multiple applications across the healthcare sector in Bangladesh:

  • Hospitals and Clinics
    Performs electrical safety testing on patient monitors, infusion pumps, ventilators, anesthesia machines, imaging equipment, and other electrically powered devices.

  • Diagnostic and Specialty Centers
    Ensures compliance for devices in cardiology, dialysis, laboratory, and imaging departments.

  • Biomedical Service Providers
    Portable design allows on-site testing across multiple facilities, enabling service companies to maintain high safety standards.

  • Medical Equipment Workshops
    Supports pre-delivery inspections, post-repair verification, and preventive maintenance programs for high-value medical equipment.

Regular use of the ULT800 reduces electrical hazards, improves equipment reliability, and enhances patient and staff safety.

Benefits for Hospitals and Biomedical Teams

The Fluke ULT800 offers several significant advantages:

  1. Advanced Safety Assurance
    Measures ground continuity, equipment leakage, and patient-applied part leakage to reduce electrical risks.

  2. Efficient and Automated Testing
    Pre-set automated sequences save time and minimize human error during routine checks.

  3. Standardization of Procedures
    Enables consistent, repeatable testing of all medical devices across departments.

  4. Support for Preventive Maintenance
    Early detection of electrical issues prevents unexpected downtime and equipment failure.

  5. Documentation and Traceability
    Digital display and data logging facilitate proper recording of test results for audits and quality assurance.

Typical Test Workflow with ULT800

A typical workflow for using the ULT800 in Bangladeshi healthcare facilities may include:

  • Visual Inspection – Examine device cables, plugs, and housing for damage.
  • Ground Wire Resistance Test – Verify protective earth continuity.
  • Equipment Leakage Current Test – Measure leakage from mains and enclosure parts.
  • Applied Part / Patient Leakage Test – Measure current through patient-connected parts if applicable.
  • Automated Test Sequence – Run standardized test programs to ensure all checks are performed consistently.
  • Result Evaluation – Compare measurements against facility or regulatory limits to determine pass/fail.
  • Documentation – Record measurements in maintenance logs or digital systems for traceability.

This workflow ensures efficient, safe, and documented testing of all medical equipment.

Considerations Before Using or Purchasing in Bangladesh

Before adopting the Fluke ULT800 Electrical Safety Analyzer, healthcare facilities should consider:

  • Device Compatibility
    Ensure the analyzer supports the medical equipment types used in the facility, including Class I, Class II, BF, and CF devices.

  • Training
    Staff should be trained on proper connections, electrical safety principles, and interpreting results accurately.

  • Calibration and Maintenance
    Schedule periodic calibration to maintain accuracy and reliability.

  • Integration into Maintenance Programs
    Include ULT800 testing in preventive maintenance, post-repair verification, and routine safety checks.

  • Documentation Practices
    Maintain logs for internal audits, compliance, and quality assurance.

The Fluke ULT800 Electrical Safety Analyzer is a professional-grade instrument that provides comprehensive, reliable, and efficient electrical safety testing for medical equipment in Bangladesh. Its combination of advanced measurement capabilities, automated sequences, and portability makes it ideal for hospitals, clinics, diagnostic centers, and biomedical service providers.

Key Features: Fluke ULT800 Electrical Safety Analyzer

The ULT800 provides a range of professional features for both workshop and field use:

  • Comprehensive electrical safety testing for medical devices
  • Ground (earth) continuity measurement to ensure protective earth integrity
  • Equipment leakage current testing including mains and enclosure leakage
  • Patient/applied part leakage testing for devices with patient connections
  • Automated test sequences for standardized testing procedures
  • High-resolution digital display for clear and precise readings
  • Portable, robust design suitable for bench-top or on-site use
  • Data logging and documentation support for maintenance records
  • Advanced interface with intuitive menu navigation

Specifications: Fluke ULT800 Electrical Safety Analyzer

Measurement Categories  CAT I, CAT II (per IEC 61010 standards)
Test Types Supported  Earth/ground bond test, insulation resistance, leakage current, point-to-point
Test Voltage for Insulation  500 VDC, 1000 VDC
Ground Bond Test Current  Up to 25 A AC
Earth/Ground Resistance Range  0.001 Ω to 2.0 Ω
Leakage Current Measurement  0.1 µA to 20 mA
Test Frequency  50/60 Hz
Measurement Accuracy  ±(1% + 5 counts)
Display Type  High-resolution color LCD
Interfaces  USB, Ethernet, Optional Wireless
Memory  Internal storage for test results and reports
Power Supply  100 to 240 V AC, 50/60 Hz
Operating Temperature  10°C to 40°C (50°F to 104°F)
Storage Temperature  -20°C to 60°C (-4°F to 140°F)
Dimensions (H x W x D)  Approx. 180 mm x 400 mm x 350 mm
Weight  Approx. 8.5 kg
Safety Compliance  IEC 61010-1, IEC 60601-1, IEC 62353
Applications  Medical device testing, preventive maintenance, and compliance verification
Certifications  CE, CSA, NRTL, ISO