Patient monitoring is crucial in modern healthcare, requiring accurate and reliable devices to ensure patient safety. The Fluke PS410 Patient Simulator is a high-precision tool designed to simulate vital signs, providing hospitals, clinics, and biomedical service teams in Bangladesh with a practical solution for testing, calibrating, and verifying patient monitors.
Pico Labs Limited is a reliable technology provider in Bangladesh where you can explore a wide range of Biomedical Test Equipment at the best price. From hospitals and clinics to diagnostic centers and biomedical engineering departments, you can find professional analyzers, simulators, and test tools for accurate performance checking of medical devices like infusion pumps, defibrillators, patient monitors, incubators, and ventilators.
1. Accurate Patient Monitor Testing
Simulates realistic patient conditions to verify monitor performance and accuracy.
2. Enhances Clinical Training
Provides a safe, controlled environment for hands-on training of medical staff and biomedical engineers.
3. Ensures Compliance
Supports quality assurance programs and regulatory requirements by maintaining documented test records.
4. Cost-Effective Solution
Eliminates the need for multiple devices by combining multiple simulation parameters in one compact unit.
5. Efficient Data Management
Automated storage and reporting simplify audits, preventive maintenance, and documentation.
| Parameter | Specification |
|---|---|
| Vital Signs Simulated | ECG, SpO₂, NIBP, Temperature |
| ECG Leads | 3-lead and 5-lead configurations |
| SpO₂ Range | 0 – 100% saturation, adjustable waveform |
| NIBP Range | 0 – 300 mmHg, automatic and manual modes |
| Temperature | 20°C – 42°C simulated patient temperature |
| Data Storage | Up to 5,000 simulation records |
| Display | Touchscreen with real-time waveform visualization |
| Connectivity | USB / Bluetooth |
| Power Supply | Rechargeable battery / AC adapter |
| Dimensions | 300 × 200 × 150 mm |
| Weight | 3.5 kg |
The Fluke PS410 Patient Simulator is a reliable, portable, and efficient tool for testing and calibrating patient monitors in Bangladesh. It enhances patient safety, supports regulatory compliance, and provides practical training opportunities. Its multi-parameter simulation, intuitive interface, and robust design make it an essential asset for hospitals, biomedical engineers, and training institutes.
The PS410 simulator provides essential functionalities for effective patient monitor testing and biomedical training:
ECG Simulation
Normal Sinus Rhythm: 12-lead with independent outputs referenced to RL.
Selectable Rates: 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, and 300 BPM.
Accuracy: ±1%.
Output Impedance: 940 Ω between leads.
ECG Amplitudes: 0.5, 1, and 2 mV.
Amplitude Accuracy: ±2% (Lead II).
High-Level Output: 1000x Lead II.
Waveforms: Square wave (2 Hz, 0.125 Hz), Pulse (30, 60, 120 BPM; 60 ms pulse width), Sine waves (0.5, 5, 10, 40, 50, 60 Hz at 1 mV amplitude), Triangle wave (2 Hz).
ST Segment Analysis: Elevated or depressed from -0.6 mV to 6 mV in 0.2 mV steps.
Arrhythmia Selections
Pacemaker Rhythms: Pacer non-capture, Pacer non-function, Demand pacer with occasional or frequent sinus, A-V sequential.
Artifacts: 50 Hz, 60 Hz, Muscle, Baseline, Respiration.
PVCs: Left and right ventricular focus, early, R on T, bigeminy, trigeminy, multi-focal, frequent multi-focal.
Ventricular Rhythms: PVCs at various rates, Ventricular tachycardia, Ventricular fibrillation (full and 1/2 size).
Supraventricular Rhythms: Supraventricular tachycardia, Premature atrial/nodal contraction, Asystole, Missed beat, Nodal rhythm, Irregular rhythm.
Atrial Rhythms: Atrial flutter, Atrial fibrillation (full and 1/2 size), Atrial tachycardia.
Heart Blocks: First-degree, Second-degree, Third-degree.
Bundle Branch Blocks: Right and Left bundle-branch block.
Pacemaker Simulation
Modes: Demand pacer with occasional or frequent sinus, A-V sequential.
Artifacts: 50 Hz, 60 Hz, Muscle, Baseline, Respiration.
General Specifications
Display: 2-line LCD.
Controls: 6 control keys; ON/OFF power switch.
Power: 9V battery or battery eliminator.
Housing: ABS plastic case.
Dimensions: 11.3 cm L x 7 cm W x 3.4 cm H (6.1 in L x 3.7 in W x 1.3 in H).
Weight: 0.4 kg (0.9 lb).
Operating Temperature: 15 °C to 35 °C (59 °F to 95 °F).
Storage Temperature: 0 °C to 50 °C (32 °F to 122 °F).