Patient simulators are essential tools in healthcare for testing, calibrating, and validating patient monitoring equipment. The Fluke ProSim 8 and ProSim 8P Patient Simulators provide advanced simulation of vital signs, allowing biomedical engineers and hospital teams to perform accurate and realistic testing. Renting these simulators in Bangladesh enables access to cutting-edge technology for training, preventive maintenance, and equipment verification without high capital investment. Pico Labs Limited is a reliable technology provider in Bangladesh where you can explore a wide range of Biomedical Test Equipment at the best price. From hospitals and clinics to diagnostic centers and biomedical engineering departments, you can find professional analyzers, simulators, and test tools for accurate performance checking of medical devices like infusion pumps, defibrillators, patient monitors, incubators, and ventilators.
1. Accurate Patient Monitor Testing
Simulates realistic patient conditions to verify accuracy and reliability of monitors.
2. Enhanced Training
Supports clinical training programs by creating dynamic, hands-on scenarios for students and staff.
3. Compliance and Quality Assurance
Helps hospitals meet equipment testing standards and maintain regulatory compliance.
4. Cost-Effective Rental Option
Access high-end patient simulators for short-term projects, training sessions, or preventive maintenance without purchasing.
5. Data Recording and Reporting
Automated storage of simulation data allows detailed analysis, documentation, and reporting for hospital quality control.
| Parameter | ProSim 8 / ProSim 8P Specification |
|---|---|
| Vital Signs Simulation | ECG, SpO₂, NIBP, Respiration, Temperature |
| ECG Leads | 3-lead, 5-lead, 12-lead configuration |
| SpO₂ Simulation | 0 – 100% saturation, adjustable waveform |
| NIBP Simulation | 0 – 300 mmHg, automatic and manual modes |
| Respiratory Rate | 5 – 60 breaths per minute |
| Temperature | 20°C – 42°C simulated patient temperature |
| Data Storage | Up to 10,000 simulation records |
| Display | Touchscreen with real-time waveform visualization |
| Connectivity | USB / Bluetooth / Wireless (8P model) |
| Power Supply | Rechargeable battery / AC adapter |
| Dimensions | 320 × 250 × 150 mm |
| Weight | 4.0 kg |
The Fluke ProSim 8 / ProSim 8P Patient Simulator is a versatile and reliable tool for hospitals, biomedical engineers, and training institutes in Bangladesh. Renting this simulator provides a cost-effective way to enhance patient monitor testing, clinical training, and preventive maintenance programs. Its realistic vital sign simulation, portability, and data management capabilities make it an essential asset for modern healthcare facilities and educational programs.
The ProSim 8 / 8P offers comprehensive capabilities for testing patient monitors and enhancing training programs:
| Parameter | Details |
|---|---|
| Dimensions | 14.5 × 30.2 × 8.6 cm (5.7 × 11.9 × 3.4 in) |
| Weight | 1.81 kg (4 lb) |
| Display | Color LCD |
| Battery life | ≥ 9 hrs (100 NIBP cycles) |
| Connections | USB upstream (Mini-B), USB host (Type‑A, 5 V/0.5 A) |
| Safety & EMC | IEC/EN 61010‑1, IEC 61326‑1; CE, CSA, C‑TICK, RoHS certifications |
Detailed Signal Specifications
Lead II amplitude: 0.05 – 5.0 mV (steps) with ±2% accuracy (+0.05 mV for low levels)
Other leads proportional (e.g., Lead I = 70%, V4 = 120%)
Heart rate: 10 – 360 BPM, ±1%; QRS: 80 ms adult or 40 ms pediatric
ST elevation: –0.8 to +0.8 mV
Pacemaker waveforms: amplitude 0–700 mV, width 0.1–2 ms, ±5%
Arrhythmia presets: AFib, flutter, PVCs, tachycardias, blocks, ALS rhythms etc.
ECG performance modes: waves and artifacts for testing dev.—e.g., sine, square, triangular, artifact injections
Fetal HR: fixed 60–240 BPM or IUP-synchronised; IUP waveforms with early/late decel. and accelerations, 0–90 mmHg over 90s
2 isolated channels; DC–5000 Hz exciter input; ±(1% + 1 mmHg) accuracy
Pressure range: –10 to +300 mmHg (static); dynamic curves: arterial, ventricular, PA, etc.
Rate: 0 or 10 – 150 BrPM; waveforms: normal/ventilated; impedance baseline 500–2000 Ω
Amplitude accuracy ±(5% + 0.1 Ω); includes apnea simulation
30.0–42.0 °C (0.5 °C steps); high‑accuracy presets (0/24/25/37/40 °C) with ±0.4 °C (std), ±0.1 °C (high)
Simulates thermodilution with 0 °C or 24 °C injectate; 2.5–10 L/min ±7.5% accuracy
Range: 10–400 mmHg; ProSim 8 accuracy ±(0.5% + 0.5 mmHg), 8P improved accuracy
Pre-set BP simulations for adult/neonatal; pulse & leak test supported
Range: 30%–100%; ±1–7% depending on level and UUT's R-curve; heart rate 30–300 BPM; includes artifact simulation and Rainbow compatibility.
Physical: 14.5 × 30.2 × 8.6 cm; 1.81 kg
Battery: lasts ~9 hrs; charges via AC/DC
Operating temp: specified accuracy conditions (e.g., NIBP accuracy within 10–30 °C)