Most dry-wells are calibrated by inserting a calibrated PRT and adjusting internal control sensors. However, this approach has limited value due to:
Metrology Wells overcome these issues through:
An application note is available to better understand the associated uncertainties.
Optional: For improved accuracy, Metrology Wells can be ordered with built-in electronics for external PRT reading using ITS-90 characterizations.
Fluke heat sources are among the most stable in the world—and Metrology Wells improve upon this legacy:
Only fluid baths or primary fixed-point devices offer better performance. This level of stability cannot be matched by the “off-the-shelf” controllers found in most dry-wells.
According to EA-10/13 guidelines, dry-wells should have a 40 mm zone of maximum homogeneity. Metrology Wells go further with:
Specifications for axial uniformity are published and guaranteed for each unit.
Defined as the temperature difference between wells:
Poorly designed sources or use of large probes can lead to large variations; Metrology Wells prevent this.
Loading is the temperature shift when all wells are filled versus when only one is. Metrology Wells minimize this through:
Thermal hysteresis arises mostly from internal control sensors rather than high-quality reference PRTs. It reflects the difference in measurements at the same set point approached from different directions.
Deep immersion is key to minimizing axial gradient and loading effects. Metrology Wells accommodate sensor-specific factors such as:
Well depths:
Metrology Wells are simple and intuitive to use, featuring:
Interfaces and Automation:
Pre-programmed functions (no PC required):
| Specifications | |||
| Range (at 23°C ambient) | 50°C to 700°C1 (122°F to 1292°F) | ||
| Display accuracy | ±0.2°C: 50°C to 425°C ±0.25°C: 425°C to 660°C |
||
| Stability | ±0.005°C: 50°C to 100°C ±0.01°C: 100°C to 425°C ±0.03°C: 425°C to 700°C |
||
| Axial uniformity | ±0.1°C: 50°C to 100°C ±0.25°C: 100°C to 425°C ±0.4°C: 425°C to 700°C |
||
| Radial uniformity | ±0.01°C: 50°C to 100°C ±0.025°C: 100°C to 425°C ±0.04°C: 425°C to 700°C |
||
| Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes) | ±0.02°C at 425°C ±0.04°C at 700°C |
||
| Hysteresis | 0.07°C | ||
| Well depth | 203 mm (8 in) | ||
| Resolution | 0.001°C | ||
| Display | LCD, °C or °F, user-selectable | ||
| Key pad | Ten key with decimal and ± button. Function keys, menu key, and °C / °F key. | ||
| Cooling time | 235 min: 700°C to 50°C 153 min: 700°C to 100°C |
||
| Heating time | 46 min: 50°C to 700°C | ||
| Size (H x W x D) | 366 x 203 x 323 mm (14.4 x 8 x 12.7 in) | ||
| Weight | 15 kg (33 lb) | ||
| Power | 115 V AC (±10%), or 230 V AC (±10%), 50/60 Hz, 1025 W |
||
| Computer interface | RS-232 Interface | ||
| Traceable calibration (NIST) | Data at 100°C, 200°C, 350°C, 500°C, and 660°C | ||
| 1. Calibrated to 660°C; reference thermometer recommended at higher temperatures. | |||
| Specifications | Built-in Reference Input | ||
| Temperature range | –200°C to 962°C (–328°F to 1764°F) | ||
| Resistance range | 0 Ω to 400 Ω, auto-ranging | ||
| Characterizations | ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 Callendar-Van Dusen (CVD): R0, a, b, d | ||
| Resistance accuracy | 0 Ω to 20 Ω: 0.0005 W 20 Ω to 400 Ω: 25 ppm |
||
| Temperature accuracy (does not include probe uncertainty) |
10 Ω PRTs: ±0.013°C at 0°C ±0.014°C at 155°C ±0.019°C at 425°C ±0.028°C at 700°C |
25 Ω and 100 Ω PRTs: ±0.005°C at –100°C ±0.007°C at 0°C ±0.011°C at 155°C ±0.013°C at 225°C ±0.019°C at 425°C ±0.027°C at 661°C |
|
| Resistance resolution | 0 Ω to 20 Ω: 0.0001 Ω 20 Ω to 400 Ω: 0.001 Ω |
||
| Measurement period | 1 second | ||
| Probe connection | 4-wire with shield, 5-pin DIN connector | ||
| Calibration | NVLAP accredited (built-in reference input only), NIST-traceable calibration provided | ||