The Fluke 9260 Mini Fixed-Point Cell Furnace delivers high-quality ITS-90 fixed-point calibration at a fraction of the cost of large-scale furnaces. Designed for indium, tin, zinc, and aluminum cells, the 9260 covers all primary fixed points from 156.5985 °C to 660.323 °C, using less than half the space and investment of a traditional system.
| General Specifications | |
| Temperature Range | 50 °C to 680 °C |
| Ambient Operating Range | 5 °C to 45 °C |
| Stability | ±0.03 °C to 300 °C ±0.05 °C above 300 °C |
| Vertical Gradient | Top and bottom zones adjustable by offset |
| Plateau Duration | 6–10 hours typical |
| Resolution | 0.01 ° |
| Display Scale | °C or °F, switchable |
| Immersion Depth | 229 mm (9 in) |
| Stabilization Time | 15 minutes nominal |
| Preheat Wells | 2 |
| Fault Protection | Sensor burnout and short protection, over-temperature thermal cutout |
| Display Accuracy | ±0.2 °C to 300 °C ±0.3 °C to 450 °C ±0.5 °C to 680 °C |
| Comparison Block | Two multi-hole blocks, blanks, and custom blocks available |
| Well-to-Well Gradient (in comparison block) | ±0.02 °C |
| Heating Time | 1.25 hrs. from 25 °C to 680 °C |
| Cooling Time | 10.5 hrs. from 680 °C to 100 °C |
| Comm. | RS-232 included |
| Power Requirements | 115 VAC (±10 %), 60 Hz, 11 A, or 230 VAC (±10 %), 50 Hz, 6 A, specify, 1200 W |
| Exterior Dimensions (HxWxD) | 489 x 222 x 260 mm (19.25 x 8.75 x 10.25 in) |
| Weight | 20.5 kg (45 lb.) with block |