The Fluke 9260 Mini Fixed-Point Cell Furnace delivers high-quality ITS-90 fixed-point calibration at a fraction of the cost of large-scale furnaces. Designed for indium, tin, zinc, and aluminum cells, the 9260 covers all primary fixed points from 156.5985 °C to 660.323 °C, using less than half the space and investment of a traditional system.

Key Features: Fluke 9260 Mini Fixed-Point Cell Furnace

  • Lower Total Cost – The 9260 furnace costs less than half of a large furnace, and the smaller-volume, 99.9999% pure metal cells are also much more affordable.
  • Complete ITS-90 Coverage – Use with indium, tin, zinc, and aluminum mini cells to realize all major ITS-90 fixed points required for laboratory calibration.
  • Simple, Hands-Off Operation – Place the fixed-point cell in the furnace at the end of the day. The next morning, initiate the built-in software routine for your specific cell type. Within an hour, the cell is stable and ready for precise measurements—all day long.
  • Flexible Software Control – Built-in software lets you choose between melting-point or freezing-point curves for each metal. While ITS-90 recommends freezing points, melting points are easier to realize, and the difference in uncertainty (less than 2 mK) is negligible for most labs.
  • Proven Performance – For most applications, Fluke Calibration’s mini cells at their melting points provide results indistinguishable from traditional cells at their freezing points.
  • High-Precision Comparison Calibrations – Optional comparison blocks are available for the 9260, offering a variety of pre-drilled or custom wells for high-precision calibrations at high temperatures. Each block provides a well depth of 229 mm (9 in).

Specifications: 9260 Mini Fixed-Point Cell Furnace

General Specifications
Temperature Range 50 °C to 680 °C
Ambient Operating Range 5 °C to 45 °C
Stability ±0.03 °C to 300 °C
±0.05 °C above 300 °C
Vertical Gradient Top and bottom zones adjustable by offset
Plateau Duration 6–10 hours typical
Resolution 0.01 °
Display Scale °C or °F, switchable
Immersion Depth 229 mm (9 in)
Stabilization Time 15 minutes nominal
Preheat Wells 2
Fault Protection Sensor burnout and short protection, over-temperature thermal cutout
Display Accuracy ±0.2 °C to 300 °C
±0.3 °C to 450 °C
±0.5 °C to 680 °C
Comparison Block Two multi-hole blocks, blanks, and custom blocks available
Well-to-Well Gradient (in comparison block) ±0.02 °C
Heating Time 1.25 hrs. from 25 °C to 680 °C
Cooling Time 10.5 hrs. from 680 °C to 100 °C
Comm. RS-232 included
Power Requirements 115 VAC (±10 %), 60 Hz, 11 A, or
230 VAC (±10 %), 50 Hz, 6 A, specify, 1200 W
Exterior Dimensions (HxWxD) 489 x 222 x 260 mm (19.25 x 8.75 x 10.25 in)
Weight 20.5 kg (45 lb.) with block