Fluke 725 Multifunction Process Calibrator you can easily source and outfit vis--vis any process parameter and test not far and wide-off off from any process device. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules. From the moment you pick it up, the Fluke 725 is ready to law as a powerful, yet easy-to-use ground calibrator. Its to hand controls, without menus, make operation understandable. If you throbbing to acquire it you can order not at Tools Bangladesh.
| Measurement Accuracy | ||
| Voltage DC | 30.000 V | 0.02% + 2 counts (upper display) |
| 30.000 V | 0.02% + 2 counts (lower display) | |
| 100.00 mV | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| Current DC | 24.000 mA | 0.02% + 2 counts |
| Resistance | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
| 401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
| 1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
| Frequency | 2.0 to 1000.0 CPM | 0.05% + 1 count |
| 1.0 to 1100.0 Hz | 0.05% + 1 count | |
| 1.00 to 10.00 kHz | 0.05% + 1 count | |
| Sensitivity | 1 V peak-to-peak-minimum | |
| Pressure | Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. | |
| Source Accuracy | ||
| Voltage DC | 100.00 mV | 0.02% + 2 counts |
| 10.000 V | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| Current DC | 24.000 mA (source) | 0.02% + 2 counts |
| 24.000 mA (simulate) | 0.02% + 2 counts | |
| Resistance | 15.0 to 400.0 Ω | 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
| 401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
| 1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
| Frequency | 2.0 to 1000.0 CPM | 0.05% |
| 1.0 to 1100.0 Hz | 0.05% | |
| 1.00 to 10.00 kHz | 0.25% | |
| Waveform | 5 V p-p squarewave, -0.1 V offset | |
| RTDs and Thermocouples | ||
| Measure accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.8°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Source accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| Accuracy stated for 4-wire measurement. | ||
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.4°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Specifications | ||
| Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature |
| Ramps | Slow ramp, Fast ramp, 25% step-ramp | |
| Loop power function | Voltage | 24 V |
| Accuracy | 10% | |
| Maximum current | 22 mA, short circuit protected | |
| Step functions | Source functions | Voltage, current, resistance, frequency, temperature |
| Steps | 25% of range, 100% of range | |
| Environmental Specifications | ||
| Operating temperature | -10°C to +55°C | |
| Storage temperature | -20°C to 71°C | |
| Operating altitude | 3000 m | |
| Safety Specifications | ||
| Agency approvals | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | |
| Mechanical and General Specifications | ||
| Size | 130 x 236 x 61 mm | |
| Weight | 0.65 kg | |
| Batteries | 4 AA alkaline batteries | |
| Battery replacement | Separate battery compartment, accessable without breaking calibration seal | |
| Side port connections | Pressure module connector, also used for remote real-time programming | |