Fluke 6054/6055/7007 Deep-Well Baths

The 7007 is designed for the temperature range of –5 °C to 110 °C, has built-in refrigeration, is 24 inches deep, and comes with a removable polycarbonate cover.

The 6054 covers the temperature range of 50 °C to 300 °C, is also 24 inches deep, and comes with a removable stainless steel cover. The 6055 is engineered for the temperature range of 200 °C to 550 °C with salt and is 17 inches deep. Specific size differences and various specifications are shown in the comparison table.

The Model 6055, operating up to 550 °C, uses molten salts with a pumping system for maintaining the necessary consistent fluid level required for liquid-in-glass thermometer calibrations. A viewing channel is built into the top cover for a clear visual path to your glass thermometers.

The 6055 also has an optional thermometer carousel for holding several glass thermometers in the correct calibration position without exposing them to the hot salts in the bath. The Model 2018 Carousel is completely constructed of stainless steel and has an elevated handle for rotating your thermometers to the viewing position.

These deep-well baths are built to the same performance standards as all Fluke Calibration baths, which means you can’t find another bath that has better stability or uniformity.

Key Features: Fluke 6054/6055/7007 Deep-Well Baths

  • Extra-deep wells suitable for thermometry work requiring additional tank depth and ultimate stability.
  • Constant liquid levels maintained through a concentric-tube design.
  • Special design tailored for the sighting of Liquid-In-Glass (LIG) thermometers.
  • Well depths varying from 17 to 24 inches, eliminating stem conduction effects in probes requiring deep immersion.

Key Features: Specifications: 6054/6055/7007 Deep-Well Baths

  7007 6054 6055
Range –5 °C to 110 °C 50 °C to 300 °C 200 °C to 550 °C
Temperature Stability ± 0.001 °C at 0 °C (water/ethanol)  
± 0.001 °C at 25 °C (water)
± 0.003 °C at 100 °C (oil 5012)
± 0.003 °C at 100 °C (oil 5012)
± 0.005 °C at 300 °C (oil 5017)
± 0.003 °C at 200 °C (salt)
± 0.01 °C at 550 °C (salt)
Temperature Uniformity ± 0.004 °C at 0 °C (ethanol)
± 0.007 °C at 100 °C (oil 5012)
± 0.007 °C at 100 °C (oil 5012)
± 0.015 °C at 300 °C (oil 5017)
± 0.005 °C at 200 °C (salt)
± 0.010 °C at 550 °C (salt)
Temperature Setting   Digital display with push-button data entry
Set-Point Resolution   0.002 °C, high res. 0.00003 °C 0.01 °C, high res. 0.00018 °C
Display Temperature Resolution 0.01 °C
Digital Setting Accuracy ± 1 °C
Digital Setting Repeatability ± 0.005 °C ± 0.01 °C
Heaters 250 to 1000 W 250 to 1000 W 260 to 2080 W

Working Area
178 mm dia. (7 in) 196 mm dia. (7.7 in) 107 mm dia. (4.2 in)

Depth
24 in 24 in 24 in
Wetted Parts 304 stainless steel
Power 230 V ac (± 10 %), 50 or 60 Hz, 15 A
(Specify frequency, contact Fluke Calibration if CE
mark required.)
230 V ac (± 10 %), 50/60 Hz,
15 A max
230 V ac (± 10 %), 50/60 Hz, 15 A
Volume 42 liters (11.2 gal) 50 liters (13.2 gal) 19.8 liters
(5.2 gal, 43 kg [95 lb] of bath salt)
Size (D x W x H) 470 x 775 x 194 mm to working surface (18.5 x 30.5 x 47 in), 1397 mm (55 in) to top of stir motor, 914 mm (36 in) to control panel 572 x 762 x 1219 mm to working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top of stir motor box, 914 mm (36 in) to control panel 572 x 775 x 1219 mm to working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to top of stir motor box, 914 mm (36 in) to control panel
Distance from Line of Sight to Top of Fluid 9.5 mm (3/8 in) 15.9 mm (5/8 in)
Automation  Use the 1586A Super-DAQ to automate temperature sensor calibration. For IEEE-488, add the 2001-IEEE to the automation package.

Weight
 
70.8 kg (156 lb)