The 52120A is designed for users whose ability to address their calibration workload may be
limited by the output current, accuracy, and drive capability of their present test equipment,
including:
The 52120A operates as a transconductance amplifier with:
*Special closed-loop operating mode
You may also operate your 52120A in closed-loop mode, seamlessly communicating with your Fluke Calibration 6105A or 6100B Electrical Power Standard to deliver enhanced 52120A accuracy.
| Operating limits | |
| Output ranges | 2 A, 20 A, 120 A (100 A dc) |
| Input current range | 200 mA, 200 mA, 120 mA (100 mA dc) |
| Current gain | 10, 100, 1000 |
| Input voltage range | 2 V, 2 V, 1.2 V (1.0 V dc) |
| Transconductance | 1,10,100 Siemens |
| Frequency | To 10 kHz |
| Maximum output compliance voltage | 4.5 Vrms (6.4 Vpeak) |
| Inductive drive capability: | 1 mH load |
| Output isolation, current terminal to earth | 600 Vrms, 850 Vpeak, dc to 850 Hz |
| Performance specifications | |
| AC accuracy, closed loop with 6105A | To 140 ppm |
| DC accuracy, standalone | To 160 ppm |
| AC accuracy, standalone | To 350 ppm |
| Phase angle accuracy | 0.006° |
| Load dependent phase angle shift | <0.001° @ 60 Hz |
| General specifications | |
| Input line voltage range | 100 V to 240 V with up to ±10 % fluctuations |
| Transient overvoltage | Impulse withstand (overvoltage); Category II of IEC 60364-4-443 |
| Frequency | 47 Hz to 63 Hz |
| Maximum consumption | < 1500 VA |
| Calibration Documentation | Report of calibration with data included; 17025 accredited report optional |
| Dimensions with feet (H x W x L) | 192 mm x 432 mm x 645 mm (7.6 inches x 17 inches x 25.5 inches) |
| Dimensions without feet (H x W x L) | 178 mm x 432 mm x 645 mm (7 inches x 17 inches x 25.5 inches) |
| Weight | 25 kg (54 lb.) |
| Design standards and compliance | Designed to EN 61010-1: 2010, CAN/CSA 22.2 No 61010.1-04, ANSI/UL 61010-1:2004, EN 61326-1:2006 CE marked, CSA listed |
| Operating temperature | 5 °C to 35 °C |
| Calibration temperature range | 16 °C to 30 °C |
| Warm up time | 1 hour |
| Safe operating max. relative humidity (non-condensing) | <80 % 5 °C to 31 °C ramping linearly down to 50 % at 35 °C |
| Operating altitude | 0 m to 2,500 m |
| Shock and vibration | MIL-PRF-28800F class 3 |
| Specification confidence level | 99% |