The Fluke 3130-G2M Portable Pressure Calibrator is designed for professionals who demand speed, accuracy, and convenience in pressure calibration. Combining robust pressure generation, high-precision measurement, electrical testing, and exceptional portability, the 3130-G2M delivers a complete solution for field and lab calibration tasks. Now you can rent Fluke 3130-G2M Portable Pressure Calibrator in Bangladesh, Compact and accurate solution for pressure measurement, testing & process calibration at best rental rates.
Connect the Fluke 3130-G2M Portable Pressure Calibrator to your PC to automate calibration routines, log results, and enhance traceability. With Fluke COMPASS for Pressure software, you can streamline workflows, improve data integrity, and deliver professional calibration documentation—right from the field.
Experience the convenience, accuracy, and rugged reliability of the Fluke 3130-G2M Portable Pressure Calibrator—your all-in-one solution for pressure calibration, wherever the job takes you.
| Environmental | |
| Operating temperature | -10°C to +50°C |
| Storage Temperature | -20°C to +60°C |
| Power requirements | 12 V DC (Universal ac adapter/charger supplied) |
| Battery | Internal 3800 mAh advanced NiMH pack |
| Operating Time | |
| On full charge | Approx 50 hours (Measure only or external air; no pump) |
| Using internal pump | Approx 100 calibration cycles to 300 psi |
| Physical | |
| Dimensions (L x W x D) | 15.25 x 12 x 7 in |
| Weight | ~7 kg (15 lb) |
| EMI/RFI conformance | EN61326:2006 Annex A |
| Connectors/ports | 1/8 in NPT (External supply port and test port) |
| Included accessories | Manual, NIST-traceable certificate, test leads, universal ac adapter/charger |
| Ranges | |
| Pressure (internal pump) | -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar) |
| Pressure (external air) | 0 to 2 MPa (0 to 300 psi, 0 to 20 bar) |
| mA | 0 to 24.000 mA |
| Volts | 0 to 30.000 V DC |
| Engineering units | psi, bar, mbar, kPa, MPa, kgf/cm2, mmH2O @ 4°C, mmH2O @ 20°C, cmH2O @ 4°C, cmH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C |
| Instrumental Measurement Uncertainty | |
| Pressure | 0.025% of reading ±0.01% FS |
| mA | 0.015% of reading ±0.002 mA |
| Volts | 0.015% of reading ±0.002 V |
| Temperature Effect (All Functions) | |
| No effect on accuracy on all functions from 15°C to 35°C | |
| Add ±0.002% F.S./°C for temps outside of 15°C to 35°C | |